Cảm biến đo điểm sương Easidew PRO I.S. Dew Point Transmitter for Moisture Analysis EPR-IS + 0/2P-500PG

Cảm biến đo điểm sương  Easidew PRO I.S. Dew Point Transmitter for Moisture Analysis  EPR-IS + 0/2P-500PG

Cảm biến đo điểm sương Easidew PRO I.S. Dew Point Transmitter for Moisture Analysis EPR-IS + 0/2P-500PG

The Easidew PRO I.S. dew-point sensor is designed to reliably and accurately measure dew point or moisture content in a wide variety of gas or liquid analyzer applications. The compact mechanical design minimises installation time and provides a robust and reliable transmitter for all global intrinsically safe hygrometer applications. The Easidew sensor can be used in combination with several of Michell’s moisture meters or moisture analyzers.

Applications

Moisture in natural gas production and processing

Pipeline drying

On line fiscal metering of gas

Dew-Point measurement in LNG production processing

Petrochemical process liquids

Trace humidity measurement in polymer production

Moisture analysis in CNG drying

Dew Points in I.S. areas

ATEX, IECEx, UKCA, cQPSus, TR CU Ex certified
  • Liên hệ
  • 1154
  • Thông tin sản phẩm
  • Bình luận
Product Easidew PRO I.S. for Gases Easidew PRO LQ I.S. for Liquids
Performance Specifications
Measurement Range -110 to +20°C dew point; -100 to +20°C dew point 0 to 1000 ppmW capability – factory configured to customer-required range and application
Accuracy ±1°C dew point (+20 to -60°C)
±2°C dew point (-60 to -110°C)
Response Time 5 mins to T95 (dry to wet)
Repeatability 0.5°C dew point
Calibration Traceable 13-point calibration and certificate
Electrical Specifications
Output Signal 4-20 mA (2-wire connection, current source); User configurable over range
Output Dew point or moisture content Moisture content
Analog Output Scaled Range Dew point:-110 to +20°C or –100 to +20°CMoisture content in gas:0-3000 ppmV
Non-standard:mg/m3, lbs/MMSCF natural gas
Moisture content in liquid: 0 to 1000 ppmW capability – factory configured to customer-required range and application
Supply Voltage 14 to 28 V DC
Load Resistance Max 250 Ω @ 12 V (500 Ω @ 24 V)
Current Consumption 23 mA max, depending on output signal
Saturation Constants
(for moisture in liquids measurements only)
  6-point look-up table for saturation constants up to 1000 ppmW over the temperature range 0 to +50°C; Saturation constants for 10 common liquids can be programmed into the Easidew PRO XP IS via the application software;
Alternatively the user can program saturation constants manually
CE Conformity Approved
Operating Specifications
Operating Temperature –40 to +60°C
Compensated Temperature Range -20 to +50°C
NOTE: The transmitter accuracy statement is only valid for the temperature range: -20/+50°C
Storage Temperature -40 to +60°C
Operating Pressure 45 MPa (450 barg) maximum
Flow rate 1 to 5 Nl/min mounted in standard sampling block
0 to 10 m/sec direct insertion
0.1 to 0.3L/min through Easidew sample block
0.1 to 1m/s direct insertion
Mechanical Specifications
Ingress Protection IP66 in accordance with standard BS EN 60529:1992;
NEMA 4 protection in accordance with standard NEMA 250-2003
Intrinsically safe area certificates ATEX/UKCA: II 1 G Ex ia IIC T4 Ga (–20°C to +70°C)

IECEx: Ex ia IIC T4 Ga (–20°C to +70°C)

TR CU 0Ex ia IIC T4 Ga (–20°C to +70°C)

cQPSus: IS, Class I, Division 1, Groups A, B, C & D, T4
Class I, Zone 0, AEx ia IIC T4 Gb, Ex ia IIC T4 Gb
Tamb +70°C
Russian Pattern Approval Russia (GOST-R), Kazakhstan (GOST-K)
Canadian Pressure Vessel Certification C.R.N. — 11 Canadian provinces
Oxygen service Optional: cleaned for enriched oxygen  
Housing material 316 stainless steel
Filter
(sensor protection)
Standard: Stainless steel sintered guard (for protection against fine particulate >80μm)
Optional: HDPE guard (for protection against fine particulate >10μm)
Process Connection and Material 316 stainless steel 5/8” — 18 UNF with bonded seal, G1/2” and 3/4” UNF adaptors available
(material certificate to BS EN 10204 3.1 — see ‘accessories and spare parts’ on product order codes)
Weight 1.27kg
Electrical Connections Screw terminal via female M20 x 1.5mm gland
Diagnostic conditions (factory programmed) Condition            Output
Sensor fault            23 mA
Under-range dew point  4 mA
Over-range dew point   20 mA
Approved galvanic isolators KFD2-CR-EX1.20200
KFD2-CR-EX1.30200
KFD0-CS-EX2.50P
KFD0-CS-EX1.50P
KFD2-STC4-EX1.H
MTL5040
MTL5541
MTL5041